title: Supported Targets
slug: nexus/supported-targets
createdAt: 2026-11-20T15:03:58.580Z
---

DT Nexus detects the targets below automatically in the Calibrate and Standards Validation windows. The first column is the name DT Nexus shows for each target.

| Target (as shown in DT Nexus)           | Type         | Serial number                        | Sharpness (SFR) and misregistration |
| --------------------------------------- | ------------ | ------------------------------------ | ----------------------------------- |
| DT Next Generation Target V2            | Reflective   | Printed: DT-AR- and 7 digits         | No                                  |
| FADGI 19264 Device Target               | Reflective   | QR code (A or F series)              | Yes, five sites                     |
| ColorChecker Digital SG                 | Reflective   | None                                 | No                                  |
| ISA Object-Level Target 1x              | Reflective   | Printed: M, P or B and 5 to 7 digits | Yes, one site                       |
| GoldenThread Device-Level Target        | Reflective   | Printed: D and 5 to 7 digits         | Yes, five sites                     |
| DT Nexus Transmissive Profiling Target  | Transmissive | QR code, or printed TP and 5 digits  | No                                  |
| DT Nexus Transmissive Validation Target | Transmissive | QR code, or printed TV and 5 digits  | No                                  |

When DT Nexus reads a serial number, it loads the measurements of that target. Otherwise it uses the generic reference for the target type. See [Reference Measurement Files](https://docs.digitaltransitions.com/nexus/reference-measurement-files).

The `SFR10`, `SFR50`, `Sharpening`, `Nyquist` and `Misregistration` results need a target with sharpness sites. See [Validation Results](https://docs.digitaltransitions.com/nexus/validation-results).

## Which Target Is Used for Profiling

When the image contains more than one target, DT Nexus picks one to create the profile (marked `CAL`) and uses the others for validation (marked `VAL`). A target whose serial number was read is always preferred over one whose serial was not read. Among those, the order of preference is:

1. DT Nexus Transmissive Profiling Target
2. DT Nexus Transmissive Validation Target
3. DT Next Generation Target V2
4. FADGI 19264 Device Target
5. GoldenThread Device-Level Target
6. ISA Object-Level Target 1x
7. ColorChecker Digital SG

To use a different target, click `CAL` on its row in the Calibrate window.

## GoldenThread Device-Level Target

The ISA GoldenThread Device-Level Target (DLT) is a 7 x 10 in portrait target with 30 numbered square patches and four circular markers. Its 12 neutral patches are used as the tone scale. DT Nexus measures sharpness, misregistration and PPI at five sites: the four corners and the center. It reads the serial number from the printed label.

DT Nexus can use it both to create a profile and to validate one. Keep these points in mind:

- **Reference values.** Without a measurement for its serial number, the target is scored against the generic DLT reference, which holds nominal printed Lab values, not a measurement. The PDF validation report then shows a yellow banner on its first page that begins "Reference is a nominal placeholder". Use a measurement of your own target for meaningful color results.
- **Profile quality.** When this target is the `CAL` target, the note "Profile quality may be limited by this ISA target's 30-patch layout." appears. For the best profile, use a target with more patches as the `CAL` target, and the DLT for validation.
- **Black point compensation.** `Preferences > Profiling` has a `Black Point Compensation` checkbox for this target, `ISA Device Level Target`, which is on by default.

## Older GoldenThread Object-Level Target Prints

DT Nexus also recognizes the older GoldenThread print of the ISA Object-Level Target 1x, which has no yellow border. It appears as ISA Object-Level Target 1x. This print carries no serial number or QR code, so DT Nexus always uses the generic OLT 1x reference for it.